(item no. 3316329)
PRODUCT OVERVIEW
If you want a very stable thermometer from –200 °C to 670 °C and are particular about the calibration, then look no further than our 5609 Secondary Reference PRT. Its short-term stability at the triple point of water is only ± 7 mK and its one-quarter inch (6.23mm) diameter lets you get accurate measurements in only 100 mm of immersion.
The 5608 is also a ± 7 mK probe at the triple point of water, but its one-eighth inch (3.18 mm) diameter gets you accurate measurements from –200 °C to 500 °C with only 80mm (3.1 in) of immersion. With multiple sheath length options on both of these probes, you can find the dimensions that are just right for your specific application.
The 5608 comes with a one-eighth inch (3.18 mm) diameter sheath in lengths of 9 inches or 12 inches. The 5609 comes with a one-quarter inch (6.35 mm) diameter sheath in lengths of 12 inches, 15 inches, and 20 inches; or with a 6 mm diameter in lengths of 300 mm, 400 mm, or 500 mm.
When looking for improved response time and reduced stem effect in shallow immersion, look for small diameter probes, because the measurement error called stem effect is caused by the diameter of the stem rather than the length of the stem.
Both of these probes have Inconel™ sheaths and are made using a special manufacturing process, giving them great precision over a wide temperature range. The sensors for these probes are reference-grade platinum and feature four-wire connections with less noisy measurements than two-wire counterparts.
As standard, each probe includes its resistance value at the triple point of water. If calibration is desired, you can order a NVLAP-accredited calibration from our laboratory; lab code 200348-0. On the report of calibration, you’ll get the test data and the ITS90 calibration coefficients that you can easily input into any Fluke Calibration thermometer. If you order your probe with a smart connector.
FEATURES
- 5608: –200 °C to 500 °C (80 mm minimum immersion)
- 5609: –200 °C to 670 °C (100 mm minimum immersion)
- Calibration not included, NVLAP-accredited calibration optional, lab code 200348-0
- Dry-well reference with 90 ° bend
SPECIFICATIONS
Temperature range |
5608: –200 °C to 500 °C 5609: –200 °C to 670 °C |
||||||||||||||||||||||
Nominal resistance at 0.01 °C | 100 Ω ± 0.5 Ω | ||||||||||||||||||||||
Temperature coefficient | 0.0039250 Ω/Ω/°C | ||||||||||||||||||||||
Relative Accuracy[1] |
± 0.007 °C at –200 °C ± 0.008 °C at 0 °C ± 0.020 °C at 420 °C ± 0.027 °C at 660 °C |
||||||||||||||||||||||
Short-term repeatability[2] |
± 0.007 °C at 0.010 °C ± 0.013 °C at max temp |
||||||||||||||||||||||
Drift[3] |
± 0.01 °C at 0.010 °C ± 0.02 °C at max temp |
||||||||||||||||||||||
Hysteresis | ± 0.01 °C maximum | ||||||||||||||||||||||
Sensor length | 30 mm ± 5 mm (1.2 in ± 0.2 in) | ||||||||||||||||||||||
Sensor location | 3 mm ± 1 mm from tip (0.1 in ± 0.1 in) | ||||||||||||||||||||||
Sheath material | Inconel™ 600 | ||||||||||||||||||||||
Minimum insulation resistance |
5608: 500 MΩ at 23 °C, 20 MΩ at 500 °C 5609: 500 MΩ at 23 °C, 10 MΩ at 670 °C |
||||||||||||||||||||||
Transition junction temperature range[4] | –50 ºC to 200 °C |
||||||||||||||||||||||
Transition junction dimensions | 71 mm x 12.5 mm (2.8 in x 0.49 in) | ||||||||||||||||||||||
Minimum immersion length[5] (<5 mK error) |
5608: 80 mm (3.1 in) 5609: 100 mm (3.9 in) |
||||||||||||||||||||||
Maximum immersion length | 305 mm (12 in) | ||||||||||||||||||||||
Response time[5] |
5608: 9 seconds typical 5609: 12 seconds typical |
||||||||||||||||||||||
Self heating (in 0 °C bath) |
5608: 75 mW/°C 5609: 50 mW/°C |
||||||||||||||||||||||
Lead-wire cable type | PTFE 24 AWG | ||||||||||||||||||||||
Lead-wire length | 1.8 m (6 ft) | ||||||||||||||||||||||
Lead-wire temperature range | –50 °C to 250 °C | ||||||||||||||||||||||
Calibration |
Calibration not included; NVLAP-accredited calibration optional, lab code 200348-0. Please see calibration uncertainty table and its explanation of changeable uncertainties. |
||||||||||||||||||||||
[1] Includes short-term repeatability and 100 hr drift. Calibration will add additional uncertainties. [2] Three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k=2) [3]After 100 hours at max temp, 95 % confidence (k=2) [4] Temperatures outside this range will cause irreparable damage. For best performance, transition junction should not be too hot to touch. [5] Per ASTM E 644 |
|
MODELS AND ACCESSORIES
Model Name | Description |
---|---|
5608-9-X |
Secondary Reference PRT, 9 in x 1/8 in, –200 to 500 °C, lead wire (cable) 6 ft (No calibration included. See recommended calibration options below. Contact your Fluke Calibration representative for additional calibration options.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5608-12-X |
Secondary Reference PRT, 12 in x 1/8 in, –200 to 500 °C, lead wire (cable) 6 ft (No calibration included. See recommended calibration options below. Contact your Fluke Calibration representative for additional calibration options.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5609-12-X |
Secondary Reference PRT, 12 in x 1/4 in, –200 to 670 °C, lead wire (cable) 6 ft (No calibration included. See recommended calibration options below. Contact your Fluke Calibration representative for additional calibration options.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5609-20-X |
Secondary Reference PRT, 20 in x 1/4 in, –200 to 670 °C, lead wire (cable) 6 ft (No calibration included. See recommended calibration options below. Contact your Fluke Calibration representative for additional calibration options.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5609-300-X |
Secondary Reference PRT, 300 mm x 6 mm, –200 to 670 °C, lead wire (cable) 6 ft (No calibration included. See recommended calibration options below. Contact your Fluke Calibration representative for additional calibration options.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5609-400-X |
Secondary Reference PRT, 400 mm x 6 mm, –200 to 670 °C, lead wire (cable) 6 ft (No calibration included. See recommended calibration options below. Contact your Fluke Calibration representative for additional calibration options.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5609-500-X |
Secondary Reference PRT, 500 mm x 6 mm, –200 to 670 °C, lead wire (cable) 6 ft (No calibration included. See recommended calibration options below. Contact your Fluke Calibration representative for additional calibration options.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5609-9BND |
Secondary Reference PRT, 15 in x 1/4 in, 9 in bend, –200 °C to 670 °C, lead wire (cable) 6 ft (No calibration included. See recommended calibration options below. Contact your Fluke Calibration representative for additional calibration options.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
Shipping | This Item Ships from Fremantle, Perth, Western Australia |
Please allow 2 - 10 days for your order to arrive.
Although Most Stocked Products will generally Ship Overnight. We source products from all over the world to bring you epic offers and the lowest prices. This means sometimes you have to wait a little longer to get your order but it's always worth it!
Returns are easy.
simply contact us for a returns number and send your item to our returns centre for fast processing. We'll get you a replacement or refund in a snap!
Description
(item no. 3316329)
PRODUCT OVERVIEW
If you want a very stable thermometer from –200 °C to 670 °C and are particular about the calibration, then look no further than our 5609 Secondary Reference PRT. Its short-term stability at the triple point of water is only ± 7 mK and its one-quarter inch (6.23mm) diameter lets you get accurate measurements in only 100 mm of immersion.
The 5608 is also a ± 7 mK probe at the triple point of water, but its one-eighth inch (3.18 mm) diameter gets you accurate measurements from –200 °C to 500 °C with only 80mm (3.1 in) of immersion. With multiple sheath length options on both of these probes, you can find the dimensions that are just right for your specific application.
The 5608 comes with a one-eighth inch (3.18 mm) diameter sheath in lengths of 9 inches or 12 inches. The 5609 comes with a one-quarter inch (6.35 mm) diameter sheath in lengths of 12 inches, 15 inches, and 20 inches; or with a 6 mm diameter in lengths of 300 mm, 400 mm, or 500 mm.
When looking for improved response time and reduced stem effect in shallow immersion, look for small diameter probes, because the measurement error called stem effect is caused by the diameter of the stem rather than the length of the stem.
Both of these probes have Inconel™ sheaths and are made using a special manufacturing process, giving them great precision over a wide temperature range. The sensors for these probes are reference-grade platinum and feature four-wire connections with less noisy measurements than two-wire counterparts.
As standard, each probe includes its resistance value at the triple point of water. If calibration is desired, you can order a NVLAP-accredited calibration from our laboratory; lab code 200348-0. On the report of calibration, you’ll get the test data and the ITS90 calibration coefficients that you can easily input into any Fluke Calibration thermometer. If you order your probe with a smart connector.
FEATURES
- 5608: –200 °C to 500 °C (80 mm minimum immersion)
- 5609: –200 °C to 670 °C (100 mm minimum immersion)
- Calibration not included, NVLAP-accredited calibration optional, lab code 200348-0
- Dry-well reference with 90 ° bend
SPECIFICATIONS
Temperature range |
5608: –200 °C to 500 °C 5609: –200 °C to 670 °C |
||||||||||||||||||||||
Nominal resistance at 0.01 °C | 100 Ω ± 0.5 Ω | ||||||||||||||||||||||
Temperature coefficient | 0.0039250 Ω/Ω/°C | ||||||||||||||||||||||
Relative Accuracy[1] |
± 0.007 °C at –200 °C ± 0.008 °C at 0 °C ± 0.020 °C at 420 °C ± 0.027 °C at 660 °C |
||||||||||||||||||||||
Short-term repeatability[2] |
± 0.007 °C at 0.010 °C ± 0.013 °C at max temp |
||||||||||||||||||||||
Drift[3] |
± 0.01 °C at 0.010 °C ± 0.02 °C at max temp |
||||||||||||||||||||||
Hysteresis | ± 0.01 °C maximum | ||||||||||||||||||||||
Sensor length | 30 mm ± 5 mm (1.2 in ± 0.2 in) | ||||||||||||||||||||||
Sensor location | 3 mm ± 1 mm from tip (0.1 in ± 0.1 in) | ||||||||||||||||||||||
Sheath material | Inconel™ 600 | ||||||||||||||||||||||
Minimum insulation resistance |
5608: 500 MΩ at 23 °C, 20 MΩ at 500 °C 5609: 500 MΩ at 23 °C, 10 MΩ at 670 °C |
||||||||||||||||||||||
Transition junction temperature range[4] | –50 ºC to 200 °C |
||||||||||||||||||||||
Transition junction dimensions | 71 mm x 12.5 mm (2.8 in x 0.49 in) | ||||||||||||||||||||||
Minimum immersion length[5] (<5 mK error) |
5608: 80 mm (3.1 in) 5609: 100 mm (3.9 in) |
||||||||||||||||||||||
Maximum immersion length | 305 mm (12 in) | ||||||||||||||||||||||
Response time[5] |
5608: 9 seconds typical 5609: 12 seconds typical |
||||||||||||||||||||||
Self heating (in 0 °C bath) |
5608: 75 mW/°C 5609: 50 mW/°C |
||||||||||||||||||||||
Lead-wire cable type | PTFE 24 AWG | ||||||||||||||||||||||
Lead-wire length | 1.8 m (6 ft) | ||||||||||||||||||||||
Lead-wire temperature range | –50 °C to 250 °C | ||||||||||||||||||||||
Calibration |
Calibration not included; NVLAP-accredited calibration optional, lab code 200348-0. Please see calibration uncertainty table and its explanation of changeable uncertainties. |
||||||||||||||||||||||
[1] Includes short-term repeatability and 100 hr drift. Calibration will add additional uncertainties. [2] Three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k=2) [3]After 100 hours at max temp, 95 % confidence (k=2) [4] Temperatures outside this range will cause irreparable damage. For best performance, transition junction should not be too hot to touch. [5] Per ASTM E 644 |
|
MODELS AND ACCESSORIES
Model Name | Description |
---|---|
5608-9-X |
Secondary Reference PRT, 9 in x 1/8 in, –200 to 500 °C, lead wire (cable) 6 ft (No calibration included. See recommended calibration options below. Contact your Fluke Calibration representative for additional calibration options.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5608-12-X |
Secondary Reference PRT, 12 in x 1/8 in, –200 to 500 °C, lead wire (cable) 6 ft (No calibration included. See recommended calibration options below. Contact your Fluke Calibration representative for additional calibration options.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5609-12-X |
Secondary Reference PRT, 12 in x 1/4 in, –200 to 670 °C, lead wire (cable) 6 ft (No calibration included. See recommended calibration options below. Contact your Fluke Calibration representative for additional calibration options.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5609-20-X |
Secondary Reference PRT, 20 in x 1/4 in, –200 to 670 °C, lead wire (cable) 6 ft (No calibration included. See recommended calibration options below. Contact your Fluke Calibration representative for additional calibration options.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5609-300-X |
Secondary Reference PRT, 300 mm x 6 mm, –200 to 670 °C, lead wire (cable) 6 ft (No calibration included. See recommended calibration options below. Contact your Fluke Calibration representative for additional calibration options.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5609-400-X |
Secondary Reference PRT, 400 mm x 6 mm, –200 to 670 °C, lead wire (cable) 6 ft (No calibration included. See recommended calibration options below. Contact your Fluke Calibration representative for additional calibration options.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5609-500-X |
Secondary Reference PRT, 500 mm x 6 mm, –200 to 670 °C, lead wire (cable) 6 ft (No calibration included. See recommended calibration options below. Contact your Fluke Calibration representative for additional calibration options.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
5609-9BND |
Secondary Reference PRT, 15 in x 1/4 in, 9 in bend, –200 °C to 670 °C, lead wire (cable) 6 ft (No calibration included. See recommended calibration options below. Contact your Fluke Calibration representative for additional calibration options.) X = termination. Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs). |
Shipping
Shipping | This item ships to |
Delivery & Returns
Please allow 2 - 10 days for your order to arrive.
Although Most Stocked Products will generally Ship Overnight. We source products from all over the world to bring you epic offers and the lowest prices. This means sometimes you have to wait a little longer to get your order but it's always worth it!
Returns are easy.
simply contact us for a returns number and send your item to our returns centre for fast processing. We'll get you a replacement or refund in a snap!
Reviews
6 Great reasons to buy from us:
-
30 Days Return
If your product is faulty or damaged under manufacture warranty, simply send it back to us and we'll cheerfully replace it or fix the problem for you. -
Returns are Easy
Simply contact us for a returns number and send your item to our returns centre for fast processing. We'll get you a replacement or refund in a snap! -
Best Price Guarantee
In the unlikely event that you find your item cheaper at another online store, just let us know and we'll beat the competitor's pricing hands-down. -
We guarantee your satisfaction
We insist that you love everything you buy from us. If you're unhappy for any reason whatsoever, just let us know and we'll bend over backwards to make things right again. -
100% Safe & Secure
Ordering from Laserman Technologies is 100% safe and secure so you can rest easy. Your personal details are never shared, sold or rented to anyone either. -
Best Laser Warranties in the Business
Laserman the Best Laser Level, Pipe Laser and Laser Measuring Warranties in the Business - Now thats unbeatable value !!.